An oscilloscope is a test instrument which allows you to look at the 'shape' of electrical signals by displaying a graph of voltage against time on its screen. It is like a voltmeter with the valuable extra function of showing how the voltage varies with time. A graticule with a 1cm grid enables you to take measurements of voltage and time from the screen.
The graph, usually called the trace, is drawn by a beam of electrons striking the phosphor coating of the screen making it emit light, usually green or blue. This is similar to the way a television picture is produced.
Oscilloscopes contain a vacuum tube with a cathode (negative electrode) at one end to emit electrons and an anode (positive electrode) to accelerate them so they move rapidly down the tube to the screen. This arrangement is called an electron gun. The tube also contains electrodes to deflect the electron beam up/down and left/right.
The electrons are called cathode rays because they are emitted by the cathode and this gives the oscilloscope its full name of cathode ray oscilloscope or CRO.
A dual trace oscilloscope can display two traces on the screen, allowing you to easily compare the input and output of an amplifier for example. It is well worth paying the modest extra cost to have this facility.

Precautions

  • An oscilloscope should be handled gently to protect its fragile (and expensive) vacuum tube.
  • Oscilloscopes use high voltages to create the electron beam and these remain for some time after switching off - for your own safety do not attempt to examine the inside of an oscilloscope!


Setting up an oscilloscope

Oscilloscopes are complex instruments with many controls and they require some care to set up and use successfully. It is quite easy to 'lose' the trace off the screen if controls are set wrongly! There is some variation in the arrangement and labelling of the many controls so the following instuctions may need to be adapted for your instrument.
    Oscilloscope trace
    This is what you should see
    after setting up, when there
    is no input signal connected
  1. Switch on the oscilloscope to warm up (it takes a minute or two).
  2. Do not connect the input lead at this stage.
  3. Set the AC/GND/DC switch (by the Y INPUT) to DC.
  4. Set the SWP/X-Y switch to SWP (sweep).
  5. Set Trigger Level to AUTO.
  6. Set Trigger Source to INT (internal, the y input).
  7. Set the Y AMPLIFIER to 5V/cm (a moderate value).
  8. Set the TIMEBASE to 10ms/cm (a moderate speed).
  9. Turn the timebase VARIABLE control to 1 or CAL.
  10. Adjust Y SHIFT (up/down) and X SHIFT (left/right) to give a trace across the middle of the screen, like the picture.
  11. Adjust INTENSITY (brightness) and FOCUS to give a bright, sharp trace.
  12. The oscilloscope is now ready to use!
    Connecting the input lead is described in the next section.
Further information on the controls: Timebase | Y amplifier | AC/GND/DC switch

Connecting an oscilloscope

co-axial lead
Construction of a co-axial lead
 
Oscilloscope probe
Oscilloscope lead and probes kit
Photograph © Rapid Electronics
 
The Y INPUT lead to an oscilloscope should be a co-axial lead and the diagram shows its construction. The central wire carries the signal and the screen is connected to earth (0V) to shield the signal from electrical interference (usually called noise). Most oscilloscopes have a BNC socket for the y input and the lead is connected with a push and twist action, to disconnect you need to twist and pull. Oscilloscopes used in schools may have red and black 4mm sockets so that ordinary, unscreened, 4mm plug leads can be used if necessary.
Professionals use a specially designed lead and probes kit for best results with high frequency signals and when testing high resistance circuits, but this is not essential for simpler work at audio frequencies (up to 20kHz).
An oscilloscope is connected like a voltmeter but you must be aware that the screen (black) connection of the input lead is connected to mains earth at the oscilloscope! This means it must be connected to earth or 0V on the circuit being tested.


 
Oscilloscope trace of AC
The trace of an AC signal
with the oscilloscope
controls correctly set
 

Obtaining a clear and stable trace

Once you have connected the oscilloscope to the circuit you wish to test you will need to adjust the controls to obtain a clear and stable trace on the screen:
  • The Y AMPLIFIER (VOLTS/CM) control determines the height of the trace. Choose a setting so the trace occupies at least half the screen height, but does not disappear off the screen.
  • The TIMEBASE (TIME/CM) control determines the rate at which the dot sweeps across the screen. Choose a setting so the trace shows at least one cycle of the signal across the screen.
    Note that a steady DC input signal gives a horizontal line trace for which the timebase setting is not critical.
  • The TRIGGER control is usually best left set to AUTO.
If you are using an oscilloscope for the first time it is best to start with an easy signal such as the output from an AC power pack set to about 4V.

Measuring voltage and time period

Wave properties The trace on an oscilloscope screen is a graph of voltage against time. The shape of this graph is determined by the nature of the input signal. In addition to the properties labelled on the graph, there is frequency which is the number of cycles per second.
The diagram shows a sine wave but these properties apply to any signal with a constant shape.

  • Amplitude is the maximum voltage reached by the signal.
    It is measured in volts, V.
  • Peak voltage is another name for amplitude.
  • Peak-peak voltage is twice the peak voltage (amplitude). When reading an oscilloscope trace it is usual to measure peak-peak voltage.
  • Time period is the time taken for the signal to complete one cycle.
    It is measured in seconds (s), but time periods tend to be short so milliseconds (ms) and microseconds (µs) are often used. 1ms = 0.001s and 1µs = 0.000001s.
  • Frequency is the number of cycles per second.
    It is measured in hertz (Hz), but frequencies tend to be high so kilohertz (kHz) and megahertz (MHz) are often used. 1kHz = 1000Hz and 1MHz = 1000000Hz.
    frequency  =           1             and     time period  =           1        
    time period frequency

 
Oscilloscope trace of AC
The trace of an AC signalY AMPLIFIER: 2V/cm
TIMEBASE: 5ms/cm
Example measurements:
peak-peak voltage = 8.4V
amplitude voltage = 4.2V

time period = 20ms
frequency = 50Hz

 

Voltage

Voltage is shown on the vertical y-axis and the scale is determined by the Y AMPLIFIER (VOLTS/CM) control. Usually peak-peak voltage is measured because it can be read correctly even if the position of 0V is not known. The amplitude is half the peak-peak voltage. If you wish to read the amplitude voltage directly you must check the position of 0V (normally halfway up the screen): move the AC/GND/DC switch to GND (0V) and use Y-SHIFT (up/down) to adjust the position of the trace if necessary, switch back to DC afterwards so you can see the signal again.
Voltage = distance in cm × volts/cm
Example: peak-peak voltage = 4.2cm × 2V/cm = 8.4V
amplitude (peak voltage) = ½ × peak-peak voltage = 4.2V


Time period

Time is shown on the horizontal x-axis and the scale is determined by the TIMEBASE (TIME/CM) control. The time period (often just called period) is the time for one cycle of the signal. The frequency is the number of cyles per second, frequency = 1/time period Ensure that the variable timebase control is set to 1 or CAL (calibrated) before attempting to take a time reading.
Time = distance in cm × time/cm
Example: time period = 4.0cm × 5ms/cm = 20ms
    and   frequency = 1/time period = 1/20ms = 50Hz



Oscilloscope, slow timebase
Slow timebase, no input
You can see the dot moving
Oscilloscope, fast timebase
Fast timebase, no input
The dot is too fast to see
so it appears to be a line

Timebase (time/cm) and trigger controls

The oscilloscope sweeps the electron beam across the screen from left to right at a steady speed set by the TIMEBASE control. Each setting is labelled with the time the dot takes to move 1cm, effectively it is setting the scale on the x-axis. The timebase control may be labelled TIME/CM. At slow timebase settings (such as 50ms/cm) you can see a dot moving across the screen but at faster settings (such as 1ms/cm) the dot is moving so fast that it appears to be a line.
The VARIABLE timebase control can be turned to make a fine adjustment to the speed, but it must be left at the position labelled 1 or CAL (calibrated) if you wish to take time readings from the trace drawn on the screen.
The TRIGGER controls are used to maintain a steady trace on the screen. If they are set wrongly you may see a trace drifting sideways, a confusing 'scribble' on the screen, or no trace at all! The trigger maintains a steady trace by starting the dot sweeping across the screen when the input signal reaches the same point in its cycle each time.
For straightforward use it is best to leave the trigger level set to AUTO, but if you have difficulty obtaining a steady trace try adjusting this control to set the level manually.


Y amplifier (volts/cm) control

Oscilloscope trace of varying DC
Varying DC (always positive)
 
The oscilloscope moves the trace up and down in proportion to the voltage at the Y INPUT and the setting of the Y AMPLIFIER control. This control sets the voltage represented by each centimetre (cm) on the the screen, effectively it is setting the scale on the y-axis. Positive voltages make the trace move up, negative voltages make it move down. The y amplifier control may be labelled Y-GAIN or VOLTS/CM.
The input voltage moving the dot up and down at the same time as the dot is swept across the screen means that the trace on the screen is a graph of voltage (y-axis) against time (x-axis) for the input signal.


The AC/GND/DC switch

Oscilloscope, input 0V
Switching to GND allows you
to quickly check the position
of 0V (normally halfway up).
The normal setting for this switch is DC for all signals, including AC! Switching to GND (ground) connects the y input to 0V and allows you to quickly check the position of 0V on the screen (normally halfway up). There is no need to disconnect the input lead while you do this because it is disconnected internally.
Switching to AC inserts a capacitor in series with the input to block out any DC signal present and pass only AC signals. This is used to examine signals showing a small variation around one constant value, such as the ripple on the output of a smooth DC supply. Reducing the VOLTS/CM to see more detail of the ripple would normally take the trace off the screen! The AC setting removes the constant (DC) part of the signal, allowing you to view just the varying (AC) part which can now be examined more closely by reducing the VOLTS/CM. This is shown in the diagrams below:
Displaying a ripple signal using the AC switch
Ripple signal Ripple signal Ripple signal
Switch in normal DC position.
The ripple is difficult to see,
but if VOLTS/CM is reduced
to enlarge it the trace will
disappear off the screen!
Switch moved to AC position.
The constant (DC) part of the
signal is removed, leaving
just the ripple (AC) part.
VOLTS/CM reduced to
enlarge the ripple.

The ripple can now be
examined more closely.